Calibrating Gauges 10
FastCalXP Operation Manual
3 Select a model from the pull-down list. If the model of the gauge you are creating has not been previously entered into the database, click on the Add New
Model button. The Add New Model Wizard will step you through the creation of a new model in two steps.
The initial dialog bo
x allows you to enter the model information.
4 Enter the Model number or name and select the Units of measure from the pull-down list.
5 Enter the Pressure range measured by the model.
6 Specify the increment represented by the Resolution of the gauge or transmitter.
When Use custom certi c
a
te template is checked, you can override the default Excel template de ned in the FastCalXP Options dialog. This allows this
model to have a unique certi cate format.
7 Enter the Cal Interval, in days, which will be used for calibration scheduling.
If you will be calibrating a transmitter, check the Transmitter box and select the parameters that you wish to measure. If you wish to Measure mA, check
the box, enter the Low and High PSI represented by the range of mA readings for the transmitter. In addition, it is possible to Measure local display or
Measure HART signals through a connected HART modem.
Note: The default certi cat
e template will change depending on the chosen transmitter con guration. Also, if Measure HART is selected, the Resolution will
automatically be adjusted to the resolution of the connected pressure standard.